Comparison of Microscan and MALDI-TOF-Mass Spectrometry Identification Methods for E. coli Isolates from Natural Waters in Kentucky, USA

Grade Level at Time of Presentation

Senior

Major

Biomedical Sciences, Pre-Medical Concentration

Minor

Chemistry

Institution 24-25

Eastern Kentucky University

KY House District #

94

KY Senate District #

31

Department

Department of Environmental & Public Health

Abstract

Introduction: E. coli are fecal indicator bacteria and their density in the environment can be indicative of fecal-oral pathogen diseases. MALDI-TOF-mass spectrometry (MS) improves bacterial identification speed and accuracy, while Microscan is known for classifying and assessing antibiotic susceptibility in clinical isolates. This study compares both methods using environmental E. coli isolates to explore their potential.

Methods: 113 water samples were collected over 13 weeks from 62 locations in Central and Eastern Kentucky. Bacterial colonies were isolated from samples using modified mTEC agar plates from presumably positive E. coli containing wells from Idexx Colilert and ColiGlow methods. 385 isolates were obtained and re-evaluated for their utilization of ONPG, MUG, and Magenta Glucoronide. Of the isolates, 121 were evaluated for species identification using the NUC-85 Microscan panels and the Vitek MALDI-TOF-MS Prime. Stata 15 and diagtest command were used to evaluate the diagnostic statistics of the Microscan for classifying E. coli as determined by the MALDI-TOF-MS.

Results/Discussion: Of the 121 isolates, 108 (89%) were deemed E. coli by the MALDI-TOF-MS, though two were not identified. Microscan agreed with 99 (92% [sensitivity]) of 108 E. coli-determined identifications, and 9 (82% [specificity]) of 11 non-E. coli identifications. Among the five (45%) of 11 discordant identifications, Microscan had low-probability identifications. Notably, some environmental isolates of E. coli (nearly 8%) were classified by the Microscan as biochemically-related species (Enterobacter spp., Klebsiella spp., Citrobacter spp., Salmonella enterica) but not E. coli, which could affect clinical diagnoses.

Conclusions: Microscan showed excellent agreement with MALDI-TOF-MS, despite relying on biochemical tests. The MALDI-TOF-MS methods, while quicker, do not yet provide a means for assessing antibiotic susceptibility reliably, and thus the Microscan remains important. Our results illustrate that the Microscan identifications for environmental isolates are quite accurate and can be coupled with antibiotic susceptibility profiles from environmental samples for informing public health surveillance efforts.

This document is currently not available here.

Share

COinS
 

Comparison of Microscan and MALDI-TOF-Mass Spectrometry Identification Methods for E. coli Isolates from Natural Waters in Kentucky, USA

Introduction: E. coli are fecal indicator bacteria and their density in the environment can be indicative of fecal-oral pathogen diseases. MALDI-TOF-mass spectrometry (MS) improves bacterial identification speed and accuracy, while Microscan is known for classifying and assessing antibiotic susceptibility in clinical isolates. This study compares both methods using environmental E. coli isolates to explore their potential.

Methods: 113 water samples were collected over 13 weeks from 62 locations in Central and Eastern Kentucky. Bacterial colonies were isolated from samples using modified mTEC agar plates from presumably positive E. coli containing wells from Idexx Colilert and ColiGlow methods. 385 isolates were obtained and re-evaluated for their utilization of ONPG, MUG, and Magenta Glucoronide. Of the isolates, 121 were evaluated for species identification using the NUC-85 Microscan panels and the Vitek MALDI-TOF-MS Prime. Stata 15 and diagtest command were used to evaluate the diagnostic statistics of the Microscan for classifying E. coli as determined by the MALDI-TOF-MS.

Results/Discussion: Of the 121 isolates, 108 (89%) were deemed E. coli by the MALDI-TOF-MS, though two were not identified. Microscan agreed with 99 (92% [sensitivity]) of 108 E. coli-determined identifications, and 9 (82% [specificity]) of 11 non-E. coli identifications. Among the five (45%) of 11 discordant identifications, Microscan had low-probability identifications. Notably, some environmental isolates of E. coli (nearly 8%) were classified by the Microscan as biochemically-related species (Enterobacter spp., Klebsiella spp., Citrobacter spp., Salmonella enterica) but not E. coli, which could affect clinical diagnoses.

Conclusions: Microscan showed excellent agreement with MALDI-TOF-MS, despite relying on biochemical tests. The MALDI-TOF-MS methods, while quicker, do not yet provide a means for assessing antibiotic susceptibility reliably, and thus the Microscan remains important. Our results illustrate that the Microscan identifications for environmental isolates are quite accurate and can be coupled with antibiotic susceptibility profiles from environmental samples for informing public health surveillance efforts.